The status register (0xE000) can report that one of the following notification types is present:
0
No issues
1
Info
2
Warning
3
Error

Additional details about the status reported in the register 0xE000 can be provided in the 16-bit register 0xE001, as described in the following table.

Bits in register 0xE001
Bit in E001hex Meaning Notes
0 (0x0000) No errors
2 (0x0002) Probe error Connect to Insight PC software and review the error information in the Diagnostics menu. Check display messages.
4 (0x0004) Low supply voltage The supply voltage is not within the specified range (15–30 V DC (24 V AC ±10 % 50/60 Hz)
16 (0x0010) No measurement probe Probe not connected or Indigo cannot detect the probe.
32 (0x0020) Probe warning Connect to Insight PC software and review the error information in the Diagnostics menu. Check display messages.