Calibration overview - INDIGO200SERIES - INDIGO202

Indigo202 User Guide

Document code
M212843EN
Revision
B
Language
English
Product
INDIGO200SERIES
INDIGO202
Content type
Product description
Product description > Features
Product description > Functions
Document type
User guide
Figure 1. Calibration menu main view example (CO2 probe)

The Calibrate menu in Insight PC software contains options for calibrating and adjusting the measurement of the probe you have connected to the transmitter. You can also view the current adjustment and restore the probe's factory adjustments.

The range of available options (for example, environmental compensations and the number of adjustment points) varies depending on the features of the connected probe.

The menu examples presented here are examples of calibration options that are available when a CO2 probe is connected to the transmitter.

CAUTION Before adjusting a probe's measurement, make sure that you have familiarized yourself with the probe-specific calibration requirements such as possible adjustment limits and environmental compensation interdependencies. See the user documentation of the probe for probe-specific information.

Starting calibration

To open the calibration menu, select to access Insight main menu. In the main menu, select Calibrate.
Figure 2. Calibration menu selection

Restoring factory adjustment

Before adjusting the probe measurement, remove any possible earlier adjustments with the Restore factory adjustment selection.
Figure 3. Restore factory adjustment selection
Always restore factory adjustment before entering a new adjustment. This prevents any possible earlier adjustments having an effect on the new adjustment you make.