Display messages - INDIGO200 - INDIGO200SERIES - INDIGO202

Indigo202 User Guide

Document code
M212843EN
Revision
B
Language
English
Product
INDIGO200SERIES
INDIGO202
Content type
Product description
Product description > Features
Product description > Functions
Document type
User guide

The following table lists the display messages that Indigo 202 uses to inform you about the transmitter's state.

In addition to the Indigo transmitter messages, the connected probes have probe-specific messages that are also shown on the display. Messages from the connected probe start with Probe:. For more information on the probe-specific messages, see the probe's documentation.

Table 1. Indigo transmitter messages shown on display
Display message Description Recommended action
Errors
Low supply voltage The supply voltage is below the minimum (range: 15 ... 30 VDC or 24 VAC ±10 % 50/60 Hz). The error clears when the supply voltage returns to the specified range. Check your power supply and restart Indigo if needed.
No legal measurement probe The connected probe is not compatible with Indigo or a restriction (probe name or serial number) on allowed probes is in place. Check your probe's Indigo compatibility and possible restrictions on allowed probes.
Warnings
No measurement probe The measurement probe is not connected, or the probe is connected poorly and cannot be recognized by Indigo. Check that you have fastened the probe correctly with the locking wheel.
Notifications
Probe connected: Indigo displays the name of the connected probe.
Waiting for measurements The connected probe is preparing measurements (for example, at start-up or after a probe restart).
Unable to measure all parameters Some of the measurements you have configured to be displayed are currently not available. This can be caused, for example, by the probe heating feature: parameters that are affected by heating cannot be shown while the heating cycle is ongoing.