Notification and error bits - INDIGO202 - INDIGO200SERIES

Indigo202 User Guide

Document code
M212843EN
Revision
A
Language
English (United States)
Product
INDIGO202
INDIGO200SERIES
Content type
Product description > Functions
Product description > Features
Product description
Document type
User guide
The status register (E000hex) can report that one of the following notification types is present:
0
No issues
1
Info
2
Warning
3
Error

Additional details about the status reported in the register E000hex can be provided in the 16-bit register E001hex, as described in the following table.

Table 1. Bits in register E001hex
Bit in E001hex Meaning Notes
0 (0000hex) No errors
2 (0002hex) Probe error Connect to Insight PC software and review the error information in the Diagnostics menu. Check display messages.
4 (0004hex) Low supply voltage The supply voltage is not within the specified range (15 ... 30 VDC (24 VAC ±10 % 50/60 Hz)
16 (0010hex) No measurement probe Probe not connected or Indigo cannot detect the probe.
32 (0020hex) Probe warning Connect to Insight PC software and review the error information in the Diagnostics menu. Check display messages.