If the analog output of the transmitter appears to be inaccurate, you can test and adjust the output level in the Insight PC software calibration menu. In analog output adjustment, the output of the transmitter's analog output channels is forced to a fixed current or voltage level. The test output can then be verified with a multimeter, and the analog output level adjusted as necessary.

For instructions on testing and adjusting analog output, see Analog output adjustment.