The transmitter has 3 analog outputs that can
be configured with the Insight PC software. Each output has the same set of configuration
options in the analog output 1 — 3 menus under the Configure Device menu (Analog output 1 shown in example).
The available selections
depend on the probe that is connected to the transmitter (for example, CO2 output
parameter selection is only available for CO2 probes).
The Analog output type selection in the General menu defines whether the outputs are used
for voltage or current output.
Receiving analog output settings
automatically from probe
The transmitter can receive the analog
output settings automatically from the connected probe in the following cases:
The transmitter is a new unit and has
not been configured yet.
The user has removed the analog output
configuration from the transmitter.
It is recommended to always
review the analog output settings with the Insight PC software after connecting a
probe.