Climet CI-3100 Modbus viewLinc system Integration Guide

Document code
M212824EN
Revision
A
Language
English
Product
viewLinc 5.1
Document type
Technical note

Most viewLinc systems will not have units for particle count or pump flow in percentage. These units will need to be created in viewLinc before you attempt to add the particle counter as a device in Device Manager.

  1. In System Preferences select the Units tab.
  2. Select Count per Cubic Foot, then select Edit.
  3. Update the unit properties:
    Name
    For non-viewLinc default unit types, you can modify the name of the unit type.
    Device Units
    Type a maximum of 4 CAPITAL letters per unit of measurement. Use a comma (,) to separate unit types.
    • Metric units:
      • PCM for particles per cubic meter
      • or C/CM for counts per cubic meter
    • Non-metric units:
      • PCF for particles per cubic foot
      • or C/CF for counts per cubic foot
    Display Text
    Type the format to display for each unit. For example, if a temperature unit is C, you may want to display Celsius.
    Min
    Set the minimum value permitted for this unit to zero.
    Max
    Set the maximum value permitted for this unit.
    Maximum value should set to the highest value, which is 2 294 967 295. This is the maximum value for a 32-bit signed integer. In practice, the maximum particle count value you will see on a Climet CI-3100 is 35,000,000 C/CM and 1,000,000 C/CF.
    Decimal Places
    Set the number of decimal place value to zero.
  4. Select Save.
  5. Perform the same steps above for Count per Cubic Meter.