Refer to your particle counter’s operator’s manual on how to connect your device.
Once you log into the particle counter web user interface, you will see the Status page. This page contains all the information you will need to configure viewLinc.
Unit ID | The Modbus TCP Unit ID for the particle counter. |
Sample Volume | Set to 1 CF, which means 1 cubic foot of air is moving through the counter every minute. |
CH 2 Rolling Counts | The live reading for 5.0um particle count that will be read by viewLinc. |
CH 1 Rolling Counts | The live reading for 0.5um particle count that will be read by viewLinc. |
Flow Value (Current) | The percentage (%) flow status that will be read by viewLinc. This verifies that 1 cubic foot of air is flowing through the particle counter. |
IP Address | The Modbus TCP IP address for the particle counter. |