| CAUTION Updating RFL100 firmware
erases the recorded data on the device. Before updating, verify from the viewLinc Enterprise
Server that up-to-date data from this device is available. You can also copy the data from
the device for local storage; see Downloading data using service port. |
Updating the firmware will not affect the
data logger's status in the Vaisala viewLinc Monitoring System. If the device was accepted
in the system before the update, it will remain accepted.
-
If the firmware update file and release notes document are in a zip package, unzip it
into a temporary directory.
-
Before starting the firmware update, read the release notes for a summary of important
changes and possible compatibility requirements.
-
If the RFL100 is switched off, switch it on.
-
Open the plug that covers the service port and connect the USB cable between your
computer and the service port of the RFL100.
After the computer detects the RFL100 and installs the appropriate
driver, it is available for file transfer.
-
Copy the firmware update file (file name ending in .fw) into the \Data\Update folder on the
RFL100. For example, if your computer has a Windows operating system, you can simply drag
the file into the correct folder (drag and drop) or use the copy and paste functions.
Select to overwrite the old file when prompted.
If the file is valid, RFL100 begins
the update automatically. Do not unplug the cable or turn off the RFL100 during the
update.

-
Monitor the update progress on the display of the RFL100. When the update is done, you
will see UPD OK message on the display, after which
the RFL100 restarts with the new firmware.
-
Disconnect the USB cable and close the plug over the service port.