Height normalization - CL51

CL51 Technical Description

Product status
Document code
M212735EN-A
ft:locale
en-US
Product
CL51
Document type
Technical description

Assuming a clear atmosphere, the power of the return signal (Pr) is inversely proportional to the square of the distance. It means that the signal strength from 1 000 ft is generally hundred times stronger than from l0 000 ft.

The height dependence is eliminated by multiplying the measured signal value with the square of height (height normalization). Noise, however, being height-independent from a measurement point of view, is then correspondingly accentuated with increasing height.