Practical measurement signal - CL51

CL51 Technical Description

Product status
Document code
M212735EN-A
ft:locale
en-US
Product
CL51
Document type
Technical description

Generally, particles at all heights scatter light, so the actual return signal may look like this:

Typical measurement signal

The instantaneous magnitude of the return signal provides information on the backscatter properties of the atmosphere at a certain height. Information about fog and precipitation, as well as clouds, can be derived from the return signal. Fog and precipitation attenuate the return signal from a cloud. However, fog and precipitation backscatter signals also provide data that can be used in estimating this attenuation and computing the necessary compensation, up to a limit.

In its normal full-range operation, CL51 digitally samples the return signal from 0 to 100 µs. The sampling is repeated every 67 ns, providing a spatial resolution of 10 meters respectively from ground level up to 15 000 meters (50 000 ft). This resolution is adequate for measuring the atmosphere, since visibility in the densest clouds is of the same order. Note that the specified cloud height reporting range is 13 km (43 000 ft).