Creating alert rules - Jade Smart Cloud

Jade Smart Cloud User Guide

Product status
Document code
M212670EN
Revision
A
ft:locale
en-US
Product
Jade Smart Cloud
Document type
User guide

Administrator

  1. Select Alert rules > Create alert rule.
  2. Enter a descriptive name for the rule.
  3. Select the measurement parameter for the rule.
  4. Select the activation limits and enter the limit values.

    You can select a High limit, a Low limit, or use both limits to alert when the value of the selected measurement parameter moves outside both limits.

    Example rule for a temperature alert with a high limit
  5. Jade automatically defines an Alert deactivation margin to prevent the alert from being repeatedly triggered when the value moves near the limit. You can adjust the margin or even remove it by setting it to zero.
  6. You can also set an Alert activation delay to prevent the alert from being activated by very short changes. By default, there is no activation delay.
    Example rule for an RH alert with high and low limits and an activation delay
  7. Select Save when done.