In pressurized environments, the actual process pressure value must be set for DM70. The pressure can be given in the following units:
- Pbarg: Gauge pressure is given in the unit of bar. It indicates the pressure difference between the normal atmospheric pressure and the actual process pressure.
- Pbara: Absolute process pressure is given in the unit of bar.
- Ppsig: Gauge pressure is given in the unit of psi. It indicates the pressure difference between the normal atmospheric pressure and the actual process pressure.
- Ppsia: Absolute pressure is given in the unit of psi.
To set the pressure values: