Autocalibration - DM70

DM70 Quick Guide

Document code
M212551EN
Revision
B
Language
English
Product
DM70
Document type
User guide

To obtain the best possible accuracy in measurements taken in dry environments, DM70 has built-in automatic calibration. During autocalibration, DM70 adjusts the dry end reading to correspond to the calibrated values.

By default, the automatic autocalibration in DM70 is turned on. In this mode, the calibration takes place automatically if the dew point or temperature changes significantly, typically more than 10 °C. However, if there are no changes in the conditions, the calibration will take place at an interval of 1 hour (or maximum 1 hour after the previous autocalibration).

If the automatic autocalibration is turned off, the autocalibration should be started when starting measuring after the probe has not been used for a while and always at least once every hour. You can start autocalibration manually in the Functions menu of the MI70 indicator.

For more information on autocalibration, see DM70 User Guide (M010091EN).