Register number (decimal) | Address (hexadecimal) | Register description | Data format | Unit / Valid range |
---|---|---|---|---|
Environmental compensation | ||||
769 | 0300hex | Power-up value for pressure compensation | 32-bit float |
hPa Range: 450 … 1550 Init/default: 1013.25 |
771 | 0302hex | Power-up value for temperature compensation | 32-bit float |
°C Range: -10 … +70 Init/default: 25 |
773 | 0304hex | Volatile pressure compensation (value cleared at probe reset) | 32-bit float |
hPa Range 450–1550 hPa Init/default: 1013.25 |
775 | 0306hex | Volatile temperature compensation (value cleared at probe reset) | 32-bit float |
°C Range: -10 … +70 Init/default: 25 |
Function control | ||||
1281 | 0500hex | Pressure compensation mode selection | Enum |
0 = Off 1 = Setpoint 2 = External |
1282 | 0501hex | Temperature compensation mode selection | Enum |
0 = Off 1 = Setpoint 2 = Measured 3 = External |
Communication | ||||
1537 | 0600hex | Serial address | 16-bit integer |
Valid range 1–255 Default: 240 |
1538 | 0601hex | Bit rate | Enum | Valid range 4800–115200 4 = 4800 5 = 9600 6 = 19200 7 = 38400 8 = 57600 9 = 115200 (default: 6 (19200)) |
1539 | 0602hex | Parity, data, stop bits | Enum |
0 = N,8,1 1 = N,8,2 2 = E,8,1 3 = E,8,2 4 = O,8,1 5 = O,8,2 |
1540 | 0603hex | Response delay | 16-bit integer | Valid range 0–1000 |
1541 | 0604hex | Restart device | Function | |
Analog output 1 | ||||
1793 | 0700hex | Analog output 1 mode selection | Enum |
0 = Disabled 1 = 4–20 mA |
1794 | 0701hex | Analog output 1 measurement parameter selection. When written, scalings are reset. | Reg |
0000hex (CH4 wet basis) 0002hex (CH4 dry basis) 1 0004hex (CO2 wet basis) 0006hex (CO2 dry basis) 1 0008hex (H2O wet basis) 1 000Ahex (H2O dry basis) 1 000Chex (Td) 1 000Ehex (Tdf) 1 See Table 1 |
1795 | 0702hex |
Scale low end for analog output 1 measurement parameter. Minimum and maximum values vary for different parameters. |
Float |
Output parameters 0000hex (CH4 wet basis) 0002hex (CH4 dry basis) 1 0004hex (CO2 wet basis) 0006hex (CO2 dry basis) 1: Minimum: 0 Maximum: 1000000 |
Output parameters 0008hex (H2O wet basis) 1 000Ahex (H2O dry basis) 1: Minimum: 0 Maximum: 250000 |
||||
Output parameters 000Chex (Td) 1 000Ehex (Tdf) 1: Minimum: -10 Maximum: 60 |
||||
1797 | 0704hex |
Scale high end for analog output 1 measurement parameter. Minimum and maximum values vary for different parameters. |
Float |
Output parameters 0000hex (CH4 wet basis) 0002hex (CH4 dry basis) 1 0004hex (CO2 wet basis) 0006hex (CO2 dry basis) 1: Minimum: 0 Maximum: 1000000 |
Output parameters 0008hex (H2O wet basis) 1 000Ahex (H2O dry basis) 1: Minimum: 0 Maximum: 250000 |
||||
Output parameters 000Chex (Td) 1 000Ehex (Tdf) 1: Minimum: -10 Maximum: 60 |
||||
1799 | 0706hex | Error output level (mA) | Float |
Min. 0.5 mA Max. 24 mA |
1801 | 0708hex | Low clipping limit | Float | 0–2.5 % |
1803 | 070Ahex | Low error limit | Float | 0–1025 % |
1805 | 070Chex | High clipping limit | Float | 0–12.5 % |
1807 | 070Ehex | High error limit | Float | 0–1025 % |
Analog output 2 | ||||
2049 | 0800hex | Analog output 2 mode selection | Enum |
0 = Disabled 1 = 4–20 mA |
2050 | 0801hex | Analog output 2 measurement parameter selection. When written, scalings are reset. | Reg |
0000hex (CH4 wet basis) 0002hex (CH4 dry basis) 1 0004hex (CO2 wet basis) 0006hex (CO2 dry basis) 1 0008hex (H2O wet basis) 1 000Ahex (H2O dry basis) 1 000Chex (Td) 1 000Ehex (Tdf) 1 See Table 1 |
2051 | 0802hex |
Scale low end for analog output 2 measurement parameter. Minimum and maximum values vary for different parameters. |
Float |
Output parameters 0000hex (CH4 wet basis) 0002hex (CH4 dry basis) 1 0004hex (CO2 wet basis) 0006hex (CO2 dry basis) 1: Minimum: 0 Maximum: 1000000 |
Output parameters 0008hex (H2O wet basis) 1 000Ahex (H2O dry basis) 1: Minimum: 0 Maximum: 250000 |
||||
Output parameters 000Chex (Td) 1 000Ehex (Tdf) 1: Minimum: -10 Maximum: +60 |
||||
2053 | 0804hex |
Scale high end for analog output 2 measurement parameter. Minimum and maximum values vary for different parameters. |
Float |
Output parameters 0000hex (CH4 wet basis) 0002hex (CH4 dry basis) 1 0004hex (CO2 wet basis) 0006hex (CO2 dry basis) 1: Minimum: 0 Maximum: 1000000 |
Output parameters 0008hex (H2O wet basis) 1 000Ahex (H2O dry basis) 1: Minimum: 0 Maximum: 250000 |
||||
Output parameters 000Chex (Td) 1 000Ehex (Tdf) 1: Minimum: -10 Maximum: +60 |
||||
2055 | 0806hex | Error output level (mA) | Float |
Min. 0.5 mA Max. 24 mA |
2057 | 0808hex | Low clipping limit | Float | 0–2.5 % |
2059 | 080Ahex | Low error limit | Float | 0–1025 % |
2061 | 080Chex | High clipping limit | Float | 0–12.5 % |
2063 | 080Ehex | High error limit | Float | 0–1025 % |
Analog output 3 | ||||
2305 | 0900hex | Analog output 3 mode selection | Enum |
0 = Disabled 1 = 4–20 mA |
2306 | 0901hex | Analog output 3 measurement parameter selection. When written, scalings are reset. | Reg |
0000hex (CH4 wet basis) 0002hex (CH4 dry basis) 1 0004hex (CO2 wet basis) 0006hex (CO2 dry basis) 1 0008hex (H2O wet basis) 1 000Ahex (H2O dry basis) 1 000Chex (Td) 1 000Ehex (Tdf) 1 See Table 1 |
2307 | 0902hex |
Scale low end for analog output 3 measurement parameter. Minimum and maximum values vary for different parameters. |
Float |
Output parameters 0000hex (CH4 wet basis) 0002hex (CH4 dry basis) 1 0004hex (CO2 wet basis) 0006hex (CO2 dry basis) 1: Minimum: 0 Maximum: 1000000 |
Output parameters 0008hex (H2O wet basis) 1 000Ahex (H2O dry basis) 1: Minimum: 0 Maximum: 250000 |
||||
Output parameters 000Chex (Td) 1 000Ehex (Tdf) 1: Minimum: -10 Maximum: +60 |
||||
2309 | 0904hex |
Scale high end for analog output 3 measurement parameter. Minimum and maximum values vary for different parameters. |
Float |
Output parameters 0000hex (CH4 wet basis) 0002hex (CH4 dry basis) 1 0004hex (CO2 wet basis) 0006hex (CO2 dry basis) 1: Minimum: 0 Maximum: 1000000 |
Output parameters 0008hex (H2O wet basis) 1 000Ahex (H2O dry basis) 1: Minimum: 0 Maximum: 250000 |
||||
Output parameters 000Chex (Td) 1 000Ehex (Tdf) 1: Minimum: -10 Maximum: +60 |
||||
2311 | 0906hex | Error output level (mA) | Float |
Min. 0.5 mA Max. 24 mA |
2313 | 0908hex | Low clipping limit | Float | 0–2.5 % |
2315 | 090Ahex | Low error limit | Float | 0–1025 % |
2317 | 090Chex | High clipping limit | Float | 0–12.5 % |
2319 | 090Ehex | High error limit | Float | 0–1025 % |
Analog input 1 | ||||
2562 | 0A01hex |
Analog input 1 compensation parameter. Read-only: controlled by the pressure and temperature compensation mode registers. |
Reg |
0204hex (pressure) 0206hex (temperature) |
2563 | 0A02hex |
Scale low end for analog input 1 compensation parameter. Minimum and maximum values vary for different parameters. |
Float |
0204hex (pressure) Minimum: 0 Maximum: 20 000 |
0206hex (temperature) Minimum: -200 Maximum: +400 |
||||
2565 | 0A04hex |
Scale high end for analog input 1 compensation parameter. Minimum and maximum values vary for different parameters. |
Float |
0204hex (pressure) Minimum: 0 Maximum: 20 000 |
0206hex (temperature) Minimum: -200 Maximum: +400 |
||||
2567 | 0A06hex | Input value (read-only) | Float |
1 MGP261 only.