Figure 1 shows the MGP260 lead-through options and breather plug location.
Cable glands are not provided by Vaisala.
When selecting cable glands for your application, review the applicable regional
requirements in Regional Ex certification overview.
Figure 1. Lead-through options and breather plug
1
Breather plug: do not adjust.
2
M16x1.5 wiring lead-through: used when wiring the optional Ex ia
external pressure or temperature sensor input terminal.
3
M20x1.5 wiring lead-through with removable plastic transport cover (fold
sides of transport cover together and pull out). Wiring option for analog output,
power supply input, and RS-485 terminals, or customer-specific additional wiring.
4
M20x1.5 wiring lead-through for analog output, power supply input, and RS-485
terminals.
CAUTION Seal unused lead-throughs. Use only Ex compliant seals on the MGP260
lead-throughs.