Figure 1. MGP260 probe parts (closed and opened view). Figure 1 shows the MGP260 main components with the connection box of the probe closed and opened. 1 Grounding terminal: use a ≥ 4 mm2 wire to connect to the grounding rail 2 Breather plug and lead-throughs for wiring: install cable glands as required and seal unused lead-throughs (see Lead-through options and breather plug) 3 Tightening flange of the connection box with captive Allen screws. Loosen the captive screws with a 5 mm (3/16 in) Allen key and open the connection box to access wiring terminals. 4 Tightening nut: only tighten from the tightening nut when installing (wrench size: 50 mm (1.97 in)) 5 1.5" male NPT thread: never install to any other thread type than 1.5" female NPT thread 6 Probe filter. Note that the probe filter must always be fastened for the instrument to function as intended. 7 1.5" NPT thread test plug 8 Wiring terminals for optional 4–20 mA input from external pressure or temperature sensor (Ex ia). This input is powered by the MGP260 probe. 9 Barrier separating the intrinsically safe (Ex ia) optional external sensor input terminals from the analog output, power supply input, and RS-485 terminals 10 Wiring terminals for 4–20 mA analog outputs, 18–30 V DC power supply input, and RS-485 communication 11 Measurement cuvette with optics and CARBOCAP® sensor inside the probe filter