MGP260 series probes have 3 scalable 4 … 20
mA analog output channels. Each output has the following configuration options:
- Output parameter selection
- Output scale low end and high end
- Clipping and error limits
- Error output level
The configuration of the outputs (output parameter selection and scaling) is selected when ordering the probe, and can be changed with Modbus or Vaisala Insight PC software.