Recommended installation position on pipeline - MGP261 - MGP262

MGP260 User Guide

Document code
M212504EN
Revision
C
ft:locale
en-US
Product
MGP261
MGP262
Document type
User guide
Figure 1. Recommended MGP260 installation position on pipeline.
Figure 1 shows the recommended installation position for MGP260. Install the probe in a straight run of pipeline, ≥ 5 pipe diameters downstream of the closest bend or other feature affecting gas flow, and ≥ 2 pipe diameters upstream to the next bend or similar feature. For best results, use the orientation and installation depth shown in Figure 1.
1
Recommended MGP260 probe installation position on pipeline.