Installation overview - MGP261 - MGP262

MGP260 User Guide

Document code
M212504EN
Revision
C
Language
English
Product
MGP261
MGP262
Document type
User guide

MGP260 can be installed either directly into the process (for example, into a threaded port or through a ball valve in the actual process pipeline), or using a flow-through adapter (for example, in a sampling line installation or when performing a field calibration).

For an example of installing MGP260 directly in the process through a threaded port, see Installation option: installation in threaded port.

For an example of installing MGP260 directly in the process using a ball valve, see Installation option: ball valve installation.

For an overview of the flow-through adapter installation option, see Installation option: flow-through adapter installation.

CAUTION Note that the pressure compensation input must be set up in order to receive accurate measurements. For pressure input options, see Pressure compensation input options.
Note that when installing MGP260 series probes into pipes with condensing gas, it is typical for the probe to require some time to heat up and recover from condensation. Installing MGP260 series probes into pipes with condensing gas and intermittent flow is not recommended.