| WARNING MGP260 series probes
have been designed for use in hazardous locations as specified by the product
classification. The personnel installing, using, or maintaining MGP260 probes are
responsible for determining the appropriate protection concept for the specific
application the probe is used in, and that the hazardous area classification of the probe
meets the requirements of the application. |
| WARNING If the equipment is used in a manner not specified by Vaisala, the
protection provided by the equipment may be impaired. |
MGP260 probes are certified for use in
hazardous areas as defined by the following classifications:
Table 1. MGP260 series hazardous area
classifications
Certification |
MGP260 classification |
UKEX / ATEX / IECEx
|
Ex II 1/2 (1) G Ex eb mb [ia] IIB T3 Ga/Gb -40 °C ≤ Tamb ≤ +60 °C |
| CAUTION The personnel installing,
operating, and maintaining MGP260 probes must have the required competencies for working in
the hazardous location, as defined by the applicable standards. |
For information on the standards that apply
to using MGP260 probes based on the classification of the device, see MGP260 certification
documentation and the declarations of conformity related to MGP260 probes at www.vaisala.com/declarationofconformity.