Insight main view - MGP261 - MGP262 - SWINSIGHT10

MGP260 User Guide

Document code
M212504EN
Revision
C
Language
English
Product
MGP261
MGP262
Document type
User guide
Figure 1. Insight main menu and settings
1

Select to access Insight main menu.

  • Configure Device: environmental compensation settings, analog input and output settings, Modbus configuration, error limits and general settings.
  • Export Settings: creates a text file export of the device settings.
  • Calibrate: options for calibrating and adjusting methane, carbon dioxide, and water vapor output, viewing adjustment data, and restoring factory adjustments.
  • Communication: contains a quick access selection for restarting the device.
  • Factory default settings: restores the probe back to default settings, clears any user adjustments, and restores the latest factory calibration.
  • About Device: general device information such as serial number and software version.
2
Select Settings to switch between the Basic Mode and Advanced Mode user modes, change the units of parameters (metric/non-metric), enter a factory code to access restricted functionalities, or view information about the Insight software.
3
Monitoring provides options for monitoring and recording selected parameters, and exporting the monitoring data as a CSV (comma-separated values) file.
4
Device information menu with the following tabs:
  • Measurements: measurement graph view with parameter drop-down selection.
  • Calibration information: read-only information about the latest stored calibration.
  • Diagnostics: troubleshooting and administrative information about the device status. Also includes an option to export the device error log as a text file. When contacting Vaisala support, it is recommended to include an up-to-date export of the error log with the support request.