To check status information:
- Use the command status in the CLI.
- See the status information in the netCDF file.
CL61 alerts can have up to 4 statuses.
The alert limits are based on alert configuration parameters. See CL61 configuration parameters.
Value | Description |
---|---|
0 | OK |
I | Indication |
W | At least one warning active, no alarms |
A | At least one alarm active |
The following tables list the possible status values in the order that they appear in the CLI status information response. The order and naming of items in the CLI status information response and in the netCDF file differ slightly.
Alert (CLI) | netCDF | Value | Description |
---|---|---|---|
Recent re-start occurred | Recently_started | 0 | Normal operation |
1 |
Device recently restarted |
Alert (CLI) | netCDF | Value | Description |
---|---|---|---|
Reporting destination URL | Measurement_data_destination_not_set | 0 | Reporting destination URL for output data set |
I | Reporting destination URL for output data not set | ||
- | Measurement_status | 0 | OK |
W | Measurement failure, contact Vaisala Technical Support | ||
- | Datacom_overall | 0 | OK |
Alert (CLI) | netCDF | Value | Description |
---|---|---|---|
Maintenance overall | Maintenance_overall | 0 | OK |
I | Prepare for maintenance | ||
W | Maintenance recommended | ||
A | Immediate maintenance needed | ||
Window condition | Window_condition | 0 | OK |
I |
Window slightly contaminated Prepare for window cleaning in near future. Measurement data is valid. |
||
W |
Window moderately contaminated Cleaning is recommended. Measurement data is valid. |
||
A |
Window dirty Immediate window cleaning needed. Measurement data is NOT valid. |
||
Window blocking | Window_blocking | 0 | OK |
W |
Window intermittently blocked |
||
A | Window blocked |
Alert (CLI) | netCDF | Value | Description |
---|---|---|---|
Hardware overall | Device_overall | 0 | OK |
W | At least one warning | ||
A | At least one alarm | ||
Transmitter | Transmitter_overall | 0 | OK |
W |
At least one warning Check transmitter warnings |
||
A |
At least one alarm Check transmitter warnings and alarms |
||
Receiver | Receiver_overall | 0 | OK |
W |
At least one warning Check receiver warnings |
||
A |
At least one alarm Check receiver warnings and alarms |
||
Servo drive module | Servo_drive_overall | 0 | OK |
W |
At least one warning Check servo drive module warnings |
||
A |
At least one alarm Check servo drive module warnings and alarms |
||
Optics | Optics_unit_overall | 0 | OK |
W |
At least one warning Check optics warnings |
||
A |
At least one alarm Check optics warnings and alarms |
||
Device control module | Device_controller_overall | 0 | OK |
W |
At least one warning Check device control module warnings |
||
A |
At least one alarm Check device control module warnings and alarms |
Alert (CLI) | netCDF | Value | Description |
---|---|---|---|
- | Transmitter_light_source | 0 | OK |
W | Warning | ||
A |
Transmitter laser power failure, contact Vaisala Technical Support |
||
TX light source power | Transmitter_light_source_power | 0 | OK |
W |
Transmitter laser power low, contact Vaisala Technical Support |
||
A |
Transmitter laser power very low, contact Vaisala Technical Support |
||
TX module electronics | Transmitter_electronics | 0 | OK |
A |
Laser power failure and memory failure Restart device and if issue persists, replace transmitter |
||
- | Transmitter_memory | 0 | OK |
A |
Memory failure Restart device and if issue persists, replace transmitter |
Alert (CLI) | netCDF | Value | Description |
---|---|---|---|
RX module electronics | Receiver_electronics | 0 | OK |
A |
Memory failure Restart device and if issue persists, replace receiver |
||
- | Receiver_memory | 0 | OK |
A |
Memory failure Restart device and if issue persists, replace receiver |
Alert (CLI) | netCDF | Value | Description |
---|---|---|---|
Polarizator electronics | Servo_drive_electronics | 0 | OK |
I | Polarizator not ready | ||
A | Polarizator initialization failure | ||
A |
Polarizator failure Restart device and if issue persists, replace device control module |
||
Polarizator electronics | Servo_drive_memory | 0 | OK |
A |
Memory failure Restart device and if issue persists, replace device control module |
||
- | Servo_drive_control | 0 | OK |
A |
Polarizator servo control failure Restart device and if issue persists, replace device control module |
||
- | Servo_drive_ready | 0 | OK |
W | Recent restart, waiting servo to be ready | ||
A | Polarizator servo is not ready, contact Vaisala Technical Support |
Alert (CLI) | netCDF | Value | Description |
---|---|---|---|
Tilt angle | Optics_unit_tilt_angle | 0 | OK |
W |
Tilt angle too steep |
||
Optics electronics | Optics_unit_electronics | 0 | OK |
A |
Memory failure Restart device and if issue persists, replace device control module |
||
Optics electronics | Optics_unit_accelerometer | 0 | OK |
A |
Tilt angle sensor failure Restart device |
||
Optics electronics | Optics_unit_memory | 0 | OK |
A |
Memory failure Restart device and if issue persists, replace device control module |
Alert (CLI) | netCDF | Value | Description |
---|---|---|---|
Internal temperature sensor | Device_controller_temperature | 0 | OK |
A |
Temperature sensor failure Restart device and if issue persists, replace device control module |
||
Module electronics | Device_controller_electronics | 0 | OK |
A |
Temperature sensor failure Restart device and if issue persists, replace device control module |
Status information example (CLI)
Device configuration
Recent re-start occurred 0 Normal operation
Data communication
Reporting destination URL 0 Reporting destination URL for output
data set
Maintenance
Maintenance overall 0 OK
Window condition 0 OK 99 %
Window blocking 0 OK
Hardware
Hardware overall 0 OK
Transmitter 0 OK
Receiver 0 OK
Servo drive module 0 OK
Optics 0 OK
Device control module 0 OK
Transmitter
TX Light source power 0 OK
TX module electronics 0 OK
Receiver
RX module electronics 0 OK
Servo drive module
Polarizator electronics 0 OK
Optics
Tilt Angle 0 OK 3.9 °
Optics electronics 0 OK
Device control module
Internal temperature sensor 0 OK
Module electronics 0 OK
Status information example in netCDF file
group: status {
// group attributes:
:Timestamp = "1640186884.550009";
:Device_controller_temperature = "OK: 0";
:Device_controller_electronics = "OK: 0";
:Device_controller_overall = "OK: 0";
:Optics_unit_accelerometer = "OK: 0";
:Optics_unit_electronics = "OK: 0";
:Optics_unit_overall = "OK: 0";
:Optics_unit_memory = "OK: 0";
:Optics_unit_tilt_angle = "OK: 0";
:Receiver_electronics = "OK: 0";
:Receiver_overall = "OK: 0";
:Receiver_memory = "OK: 0";
:Window_blocking = "OK: 0";
:Window_condition = "OK: 0";
:Servo_drive_electronics = "OK: 0";
:Servo_drive_overall = "OK: 0";
:Servo_drive_memory = "OK: 0";
:Servo_drive_control = "OK: 0";
:Servo_drive_ready = "OK: 0";
:Transmitter_electronics = "OK: 0";
:Transmitter_light_source = "OK: 0";
:Transmitter_light_source_power = "OK: 0";
:Transmitter_overall = "OK: 0";
:Transmitter_memory = "OK: 0";
:Maintenance_overall = "OK: 0";
:Device_overall = "OK: 0";
:Recently_started = "OK: 0";
:Measurement_status = "OK: 0";
:Datacom_overall = "OK: 0";
:Measurement_data_destination_not_set = "Reporting destination URL for output data set: 0";