Laptop with terminal software that
supports SSH (for example, freeware TeraTerm or latest version of PuTTY)
Ethernet maintenance cable
Use of the maintenance
connection is allowed only when maintenance operations are performed at qualified locations.
CL61 does not comply with EMC standards when the door is open. The door has a conductive
gasket to suppress electromagnetic radiation. When CL61 is used for normal operation, the
door must always be closed.
The maintenance connection is established
with the Ethernet maintenance cable included in the delivery, connecting the PC or laptop to
the maintenance port of the ceilometer. The Ethernet maintenance port is located at the
bottom of the measurement unit.
Push the plug into the port
and turn 90ยบ to lock in place.
Connect your laptop to the maintenance port of CL61 with the Ethernet maintenance
cable.
Maintenance port: eth0
Data port: eth1
Figure 1. Measurement unit connectors, back
of ceilometer
1
Ethernet maintenance port (eth0)
2
Ventilation plug
3
AC/DC in
4
Interface unit (control)
5
Interface unit (data)
CAUTION Remove the Ethernet
maintenance cable after setup. It is not intended for permanent installation.
Configure your computer settings to match the CL61 network settings.
From your computer Control Panel, open Network and Internet / Network and Sharing Center.
Select Change adapter
settings.
Right-click Ethernet and select Properties.
Double-click Internet Protocol Version 4 (TCP/IPv4).
Fill in IP address, Subnet mask, and Default gateway using the following computer settings.
Property
Value
IP address
172.17.0.5
Subnet mask
255.255.255.0
Default gateway
172.17.0.1
After configuring your computer settings, select OK on both pop-up windows to close them. The settings save and apply when the windows
are closed.
Open the command prompt app on your laptop.
To test the connection to CL61, type the following in command prompt: