Property | Description/value |
---|---|
Measurement range | 0 … 10 000 ppm CO2
(up to 30 000 ppm CO2 with reduced accuracy) |
Accuracy 1 | |
0 … 3000 ppm CO2 | ±40 ppm CO2 |
3000 … 10 000 ppm CO2 | ±2 % of reading |
Up to 30 000 ppm CO2 | ±3.5 % of reading |
Calibration uncertainty | |
at 2000 ppm CO2 | ±31 ppm CO2 |
at 10 000 ppm CO2 | ±105 ppm CO2 |
Long-term stability | |
0 … 3000 ppm CO2 | ±60 ppm CO2/year |
3000 … 6000 ppm CO2 | ±150 ppm CO2/year |
6000 … 10 000 ppm CO2 | ±300 ppm CO2/year |
Temperature dependence 0 … 10 000 ppm CO2 | |
With compensation, −10 … +50 °C | ±0.05 % of reading/°C |
With compensation, −40 … +60 °C | < ±0.1 % of reading/°C |
Without temperature compensation at 2000 ppm CO2 (typical) | −0.5 % of reading/°C |
Pressure dependence | |
With compensation at 0 … 10 000 ppm CO2, 500 … 1100 hPa | ±0.015 % of reading/hPa |
Without compensation (typical) | +0.15 % of reading/hPa |
Humidity dependence | |
With compensation, 0 … 10 000 ppm CO2, 0 … 100 %RH | ±0.7 % of reading (at +25 °C (+77 °F)) |
Without compensation (typical) |
+0.05 % of reading/%RH |
O2 dependence | |
With compensation, 0 … 10 000 ppm %CO2, 0 … 90 %O2 | ±0.6 % of reading (at +25 °C (+77 °F)) |
Without compensation (typical) | −0.08 % of reading/%O2 |
Start-up, warm-up, and response time | |
Start-up time at +25 °C | < 12 s |
Warm-up time for full spec. | < 2 min |
Response time (T90): | |
With standard filter |
< 1 min |
Flow-through option with > 0.1 l/min | 30 s |
With spray shield | < 3 min |
Flow rate dependence (for flow-through option) | |
< 1 l/min flow | No effect |
1 … 10 l/min flow | < 0.6 % of reading l/min |
Property | Description/value |
---|---|
Operating temperature of CO2 measurement | −40 … +60 °C (−40 … +140 °F) |
Storage temperature | −40 … +70 °C (−40 … +158 °F) |
Humidity | 0 … 100 %RH, non-condensing |
Condensation prevention | Sensor head heating when power on |
IP rating, probe body |
IP65: Dust-tight. Protected from water jets from any direction. |
Chemical tolerance (temporary exposure during cleaning) |
|
Pressure | |
Compensated | 500 … 1100 hPa |
Operating | < 1.5 bar |
Gas flow (for flow-through option) | |
Operating range | < 10 l/min |
Recommended range | 0.1 … 0.8 l/min |
Property | Description/value |
---|---|
Weight, probe | 58 g (2.05 oz) |
Connector type | M12 5-pin male |
Materials | |
Probe housing | PBT polymer |
Filter | PTFE |
Connector | Nickel‑plated brass |
Dimensions | |
Probe diameter | 25 mm (0.98 in) |
Probe length | 130 mm (5.12 in) |
Property | Description/value |
---|---|
Analog outputs |
|
Digital output | Over RS-485:
|
Operating voltage | |
With digital output in use | 12 … 30 V DC |
With voltage output in use | 12 … 30 V DC |
With current output in use | 20 … 30 V DC |
Power consumption | |
Typical (continuous operation) | 0.4 W |
Maximum | 0.5 W |
Property | Description/Value |
---|---|
EU directives and regulations |
EMC Directive (2014/30/EU) RoHS Directive (2011/65/EU) amended by 2015/863 |
EMC compatibility |
EN 61326‑1, basic electromagnetic environment |
EMC emissions |
CISPR 32 / EN 55032, Class B |
Compliance marks |
CE, RCM |
1 At 25 °C (77 °F) and 1013 hPa (incl. repeatability and
non-linearity).