Configuring SP Lite2 - AWS810

AWS810 Configuration and Maintenance Guide

Document code
M212419EN
Revision
H
ft:locale
en-US
Product
AWS810
Document type
User guide

SP Lite2 is connected to the DMU801 data management unit through the analog channel G3. To get correct radiation measurements, set the channel gain sensitivity correctly in the station parameters. If the sensitivity is not set, a warning is generated to the system warning log.

  1. In the Web UI, go to Settings > Station parameters.
  2. Locate the sr_d_1 parameter group on the parameter list.
  3. In parameter sensitivity, set the sensitivity to the correct value.
    The sensitivity differs between specific sensor units. Check the correct sensitivity from the sensor label.
  4. To save the changes and take them into use, select Save and restart script.