These instructions apply to 12 V / 24 to
28 Ah batteries.
Replace lead‑acid batteries every 5 years.
When operating in an environment where the ambient temperature is regularly over +25 °C (+77
°F), the battery may need to be replaced more often.
If a solar-powered station powers off during
the night, it means that the battery is not working and must be replaced.
| The backup battery does not
have any replaceable parts. |
| WARNING Do not replace the
battery with a battery of an incorrect type. If you do, there is a risk of
deflagration. Dispose of used batteries according to statutory regulations. |
| WARNING Do not throw the
battery into a hot oven or fire, or mechanically crush or cut the battery. Improper
disposal can cause an explosion. |
| WARNING Do not leave the
battery in an environment surrounded by extremely high temperatures. Extremely high
temperature can result in deflagration or leakage of flammable liquid or gas. |
| WARNING Make sure that the
battery is not subjected to an extremely low air pressure. Extremely low air
pressure can result in deflagration or leakage of flammable liquid or gas. |