For basic validation, perform field calibration with a calibrated multimeter and a voltage and resistance simulator. If you suspect any change based on the field calibration, send DMA801/DMA802 to Vaisala for calibration. Replace it with a calibrated spare DMA801/DMA802, or wait until Vaisala sends DMA801/DMA802 back from calibration.
- Remove DMA801 or DMA802 from DMU801.
- Send DMA801 or DMA802 to Vaisala for calibration.
-
Check the operation of the calibrated DMA801 or DMA802.
- Install the calibrated DMA801 or DMA802 into DMU801.
- Connect the sensors.
- Check that the sensor connections are working, and the data reports show data.