WARNING The system is powered from several power sources to ensure uninterrupted service. To make the system non-energized, disconnect all power sources. |
- 1
- AC (mains) circuit breaker
- 2
- External DC or sensor heating circuit breakers
- 3
- Enclosure heating circuit breaker
- 4
- BMU801 battery management unit's Power/Test button
- To switch off AC (mains) power, switch off power from the external disconnection device and then the AC (mains) circuit breaker.
- To switch off external DC power, switch off power from the external disconnection device and then the external DC circuit breaker.
- Switch off sensor and enclosure heating power from the sensor and enclosure heating circuit breakers.
- To switch off solar power and battery power, press the BMU801 Power/Test button.
- When replacing batteries, disconnect the battery cables from the batteries.
- When replacing BMU801, disconnect the battery cables from BMU801.
- When replacing solar panel, disconnect the solar panel cable from BMU801.