Installing HMP110 - AWS810

AWS810 Installation Guide

Document code
M212418EN
Revision
G
Language
English
Product
AWS810
Document type
User guide
Mount the humidity and temperature probe at the opposite end of the support arm from the solar radiation sensor.
To prevent corrosion, oxidation, and seizing, use a non‑conductive, non‑water‑soluble, non‑silicone‑based grease, such as ceramic or Teflon paste, between different kinds of metal.

DTR504 with probe installation kit includes a rain/solar radiation shield DTR504 and a plastic installation support for HMP110. Install the radiation shield on a support arm.

Figure 1. HMP110 probe with radiation shield
1
Radiation shield DTR504
2
Probe HMP110
3
Plastic nut
4
Installation support (227777)
5
Cable tie
  1. Fasten the probe to the installation support with a cable tie.
  2. Insert the support to the radiation shield and attach it by tightening the plastic nut.
  3. Attach the radiation shield on the support arm with a U‑bolt.