Aligning WMT700 - AWS810

AWS810 Installation Guide

Document code
M212418EN
Revision
G
Language
English
Product
AWS810
Document type
User guide
  • 5‑mm Allen key
  • Compass

When you are installing the sensor, pay attention to the North arrow printed on top of the sensor. The North arm is marked with an orange sticker. Align the North arm with the arm below the North arrow.

Do not remove the instrument or sensor from the mounting kit during alignment.
  1. Remove the transportation damper that protects the array and store it for future use.
  2. To align the sensor, use a compass or other similar method and rotate the sensor so that the arrow points North.
  3. Tighten the bolts of the mounting kit. Tightening torque 5 Nm.
  4. If you cannot align WMT700, measure the deviation angle from North and configure the wind direction offset error.