The following table shows the typical power consumption that can be used to estimate battery backup time for 12 V DC backup battery operation. Backup batteries do not support sensor heating power.AWS810 operating power consumption
Equipment Operating power consumption 1
Basic system (DMU801 data management unit, BMU801 battery management unit) 0.5 W
4G modem transmission (typical) 0.4 W
GOES transmission (1 minute every hour) 0.56 W
WMT700 1.5 W
WMT700 low-power option 0.7 W
R.M. Young Wind Monitor 0.06 W
WXT530 0.036 W
HMP155 0.02 W
HMP110 0.02 W
PTB330 0.36 W
BARO‑1 0.004 W
NR Lite2 0 W
SMP3, SMP6, SMP10, SMP22 0.055 W
SMP12 3.5 W
SP Lite2 0 W
SUV5 0.055 W
CSD3 < 0.0001 W
PWD22 6 W
FD70 10 W
SBS314 0 W
OTT Pluvio2 L 0.18 W
CL31 (typical, measurement unit only) 0.15 W
SnowVUE10 0.25 W
PAA‑36XW < 0.06 W
VEGAPULS C 23 < 0.485 W
VEGAPULS 61 0.09 W
QMT110 0.0005 W