Property | Description/Value |
---|---|
Measuring | |
Bucket content | 750 mm (29.53 in) |
30 l (7.93 gal) | |
Collecting area | 400 cm2 (62 in2) |
Ø 225.7 ±0.4 mm (Ø 8.89 ±0.017 in) | |
Sample interval (poll) | 1 … 60 min |
Precipitation | |
Recordable precipitation | Liquid, solid, mixed |
Recordable precipitation amount (version 400) | 750 mm (29.53 in) |
Wind | |
Maximum wind speed to retain accuracy | 33 m/s (74 mph) |
Maximum wind speed without device fault | 50 m/s (112 mph) |
Other | |
Relative humidity | 0 … 100 % (non-condensing) |
Measurement value | Units | Measuring range | Resolution | Accuracy absolute | Accuracy relative | Output delay (minutes) |
---|---|---|---|---|---|---|
Intensity RT | mm/h | 6.0 … 1800.0 | 0.6 | ±6 | ±5 % | < 1 |
Accumulated RT/NRT | mm | 0.10 … 500.0 | 0.01 | ±0.1 | ±5 % | 1 … 65 |
Property | Description/Value |
---|---|
Power | |
Supply voltage | 5.5 … 28 V DC, typically 24 V, secured against reverse polarity |
Current consumption | Typically 9.3 mA at 12 V |
Power consumption | ≤ 180 mW |
Orifice rim heating power | |
Supply voltage | 20 … 28 V DC, typically 24 V DC, secured against reverse polarity 1 |
Heating capacity | Typically 53 W at 24 V, max. 58 W |
Output delay | |
Real time | < 1 minute |
Non-real time | 5 … 65 minutes |
Interfaces | |
USB | Version 1.1 (only for service purposes, no overload protection) |
SDI‑12 | Version 1.3 |
Output measurement values |
Property | Description/Value |
---|---|
Operating temperature | −40 … +60 °C (−40 … +140 °F) |
Storage temperature | −50 … +70 °C (−58 … +158 °F) |
Temperature compensation | −25 … +45 °C (−13 … +113 °F) |
IP rating |
IP65 (resistant to salt fog) (pipe housing closed) IP63 (pipe housing open) IP67 (load cell) |
Orifice rim temperature | |
Operating range of the orifice rim heating (ambient temperature) | −40 … +20 °C (−40 … +68 °F) |
Measuring range of the orifice ring rim temperature | −20 … +40 °C (−4 … +104 °F) |
Set temperature (orifice ring rim) | +5 … +20 °C (+41 … +68 °F), factory setting: +8 °C (+46.4 °F) |
Accuracy of the set temperature | ±1 °C |
Property | Description/Value |
---|---|
Dimensions (H × Ø) | 660 × 450 mm (25.98 × 17.72 in) |
Weight (empty) | 15 kg (33 lb) |
EMC | EN 61000‑4‑2/3/4/5/6 |
Property | Description/Value |
---|---|
Base plate | Aluminum |
Collecting bucket | Polyethylene |
Bucket overlay | Polypropylene |
Pipe housing | ASA, UV resistant |
1 Galvanic separation of the power supply for the orifice rim
heating and the precipitation gauge is not necessary.
2 Used by data logger for reporting the precipitation
intensity.
3 Used by data logger for calculating the precipitation
amount.