Property | Description/Value |
---|---|
Sensor type | Vaisala BAROCAP® (silicon capacitive) |
Observation range | 500–1100 hPa |
Resolution | 0.01 hPa |
Operating temperature |
−40 … +60 °C (−40 … +140 °F) |
Property | Description/Value |
---|---|
Pressure range | 500–1100 hPa |
Accuracy | Class A |
Linearity 1 | ±0.05 hPa |
Hysteresis 1 | ±0.03 hPa |
Repeatability 1 | ±0.03 hPa |
Calibration uncertainty 2 | ±0.07 hPa |
Accuracy at +20 °C (+68 °F) 3 | ±0.10 hPa |
Total accuracy at −40 … +60 °C (−40 … +140 °F) | ±0.15 hPa |
Temperature dependence 4 | ±0.10 hPa |
Long-term stability | ±0.10 hPa/year |
Response time | 1 s |
Property | Description/Value |
---|---|
Dimensions (H × D) | 123 × 79 mm (4.84 × 3.11 in) |
Distance between 2 plug‑in module slots in DMU801 frame | 18 mm (0.71 in) |
Weight | 88 g (3.1 oz) |
Material | UL‑rated fire-resistive FR4 |
Handles | Stainless steel |
Coating | Conformally coated against humidity |
1 Defined as ±2 standard deviation limits of
endpoint non-linearity, hysteresis error, or repeatability error.
2 Defined as ± 2 standard
deviation limits of inaccuracy of the working standard including traceability to
NIST.
3 Defined as the root sum of
the squares (RSS) of endpoint non-linearity, hysteresis error, repeatability error
and calibration uncertainty at room temperature.
4 Defined as ±2 standard deviation
limits of temperature dependence over the operating temperature range.