Measuring field samples - PR-33 - PR-33-AC

PR-33-AC User Guide

Document code
M212411EN
Revision
B
Language
English
Product
PR-33
PR-33-AC
Document type
User guide

The Diagnostics page offers a possibility to measure samples for field calibration purposes. For more information, see Field calibration.

Measure a sample by clicking the Field point button on the page. After clicking the button the instrument measures ten measurement results and shows the average and deviation of the measurements. The measurement status is also shown, and if the status is not Normal operation, the point is not accepted.

Several points may be measured, and all the points display on the page until the page is reloaded. After measuring a sufficient amount of points, you can print the page.

Figure 1. Measuring field samples