Viewing refractometer status - PR-33 - PR-33-AC

PR-33-AC User Guide

Document code
M212411EN
Revision
B
Language
English
Product
PR-33
PR-33-AC
Document type
User guide

The basic information on the measurements is shown on the Main page of the instrument. For more information, see Main page. More information is shown on the Diagnostics page. For more information, see Parameters.

The measurement result is calculated from the refractive index (nD) and process temperature (T) values. Both of these values are available on the main page.

In addition to these measurements, the refractometer monitors its internal temperature and humidity, which both are available on the Diagnostics page. The internal temperature should not be above 65 °C, and the humidity should be below 60 %. High humidity is an indication of leaking seals, and high temperature may deteriorate the measurement performance and/or shorten the lifetime of the unit.