The configuration registers listed here are the most important for typical users. For more information on available configuration registers, see DMP Series User Guide (M212357EN).

Table 1. Modbus configuration data registers (writable)
Register number Address Description Data format Unit / Valid range
Compensation setpoints
769 0300hex Pressure compensation setpoint 32-bit float

Unit: hPa

Default: 1013.25 hPa

821 0334hex

Temperature compensation setpoint. If a value is written to this register, probe uses it instead of its own temperature measurement.

32-bit float Unit: °C
Communication
1537 0600hex Modbus address 16-bit integer

1 ... 247

Default: 240