| Only a member of the default viewLinc Administrators group can perform this task. |
If a DL data logger security status indicates tampered (in Sites Manager on the Hosts and Devices tab, Properties table, Security Status column), it is advised that a member of the Administrators group investigate the issue to determine the cause. For assistance, contact Vaisala Technical Support. Once the issue is identified and/or corrected (according to your company security policy) you can complete the following steps to reset the security status.
| Change to the security status does not interrupt continuous monitoring. |
- In Sites Manager select the Hosts and Devices tab.
- On the Hosts and Devices tree, select one or multiple DL data loggers (CTRL+click) you want to edit.
- Select .
- At the warning prompt, select Yes.
- At the completion prompt, select OK.