Configuring filtering factor - HMT370EX

HMT370EX User Guide

Document code
M212305EN
Revision
E
ft:locale
en-US
Product
HMT370EX
Document type
User guide
Figure 1. Filtering factor configuration in Insight PC software
1
Enable Filtering on/off to use the default filtering factor.
2
Enable Extended filtering on/off if you want to enter a custom filtering factor value.
3
After enabling the Extended filtering on/off selection, enter the custom value into the Filtering factor field.
You can apply a filtering factor to the measurement output to average out measurement output over time (for more information, see Environmental compensation and measurement settings overview). A default or custom filtering factor can be set using either the local display interface or Vaisala Insight PC software.
To set a filtering factor:
  1. If you are using the local display interface:
    1. Select Menu > Settings > Measurement settings > Filtering.
    2. To enable the default filtering factor (13 second moving average (T90)), set the Filtering on/off selection to ON.
    3. If you want to enter a custom filtering factor, set the Extended filtering on/off selection to ON, select Filtering factor and enter the custom value with the Up and Down buttons.
    4. After the final digit, pressing Select saves the pressure value.
  2. If you are using Insight PC software:
    1. Connect to Insight PC software: see Connecting to Insight PC software.
      Note that configuring these settings requires using Insight in Advanced Mode.
    2. Select > Configure Device > Filtering.
    3. To enable the default filtering factor (13 second moving average (T90)), set the Filtering on/off selection to ON and select Save.
    4. If you want to enter a custom filtering factor, set the Extended filtering on/off selection to ON, enter the custom value into the Filtering factor field, and select Save.