Calibration and adjustment overview - HMT370EX

HMT370EX User Guide

Document code
M212305EN
Revision
E
ft:locale
en-US
Product
HMT370EX
Document type
User guide
Figure 1. Insight PC software calibration view example
Figure 2. Local display calibration menu
The RH and T measurement of the probe used in the HMT370EX transmitter can be calibrated and adjusted (1-point or 2-point adjustments) using one of the following alternative configuration options:
  • Detaching the probe body from the transmitter and connecting the probe to Insight PC software (requires M12 Indigo USB Adapter cable accessory (Vaisala item USB2)).
  • Calibrating and adjusting the probe with Insight PC software when the probe is attached to the transmitter (requires M8 USB cable accessory 219690).
  • Calibrating and adjusting the probe with the local display interface of the transmitter when the probe is attached to the transmitter.