The RH and T measurement of the probe used
in the HMT370EX transmitter can be calibrated and adjusted (1-point or 2-point adjustments)
using one of the following alternative configuration options:
- Detaching the probe body from the transmitter and connecting the probe to Insight PC software (requires M12 Indigo USB Adapter cable accessory (Vaisala item USB2)).
- Calibrating and adjusting the probe with Insight PC software when the probe is attached to the transmitter (requires M8 USB cable accessory 219690).
- Calibrating and adjusting the probe with the local display interface of the transmitter when the probe is attached to the transmitter.