CAUTION
The transmitter body enclosure must not be opened in an explosion hazardous area, unless a safe work permit has been issued in accordance with the standard IEC 60079-14. Either remove the transmitter from the hazardous area before opening the enclosure, or ensure that an IEC 60079-14 compliant safe work procedure has been implemented in the hazardous area. |
CAUTION
When connecting several devices at the same time, note that your computer may not be able to supply enough power through its USB ports. Use an externally powered USB hub that can supply >2 W for each port. |
-
If you want to configure the probe by itself:
-
If you want to configure the transmitter and probe together as one unit:
- Open the Insight PC software.
- Connect the M8 USB cable (Vaisala item 219690) to a free USB port on the PC.
- Open the transmitter cover: see Mounting for an example.
- Reconnect the probe body to the transmitter cover.
- Connect the M8 USB cable (Vaisala item 219690) to the service port on inside of the transmitter cover.
- Wait for Insight to detect the probe.