Adjusting analog output level with local display interface - HMT370EX

HMT370EX User Guide

Document code
M212305EN
Revision
E
ft:locale
en-US
Product
HMT370EX
Document type
User guide
To adjust the output level of analog output channel 1 or 2 with the local display interface:
  1. Detach the probe body and open the transmitter cover.
  2. Connect a multimeter to the mA testing points located above the output screw terminals on the transmitter component board (see Figure 1).
  3. Select Menu > Calibration > Adjust outputs, and then select either Analog output 1 or Analog output 2.
  4. The display shows information about the analog output level adjustment. Select Next to start sending the low level (4 mA) test output from the transmitter analog channel.
  5. Verify that the output reading of the multimeter matches 4 mA. If the multimeter reading differs from 4 mA, enter the required change to the transmitter output level with the Up and Down buttons. Select Next after the correction.
  6. The transmitter starts to send the high level (20 mA) test output. Verify that the output reading of the multimeter matches 20 mA. If the multimeter reading differs from 20 mA, enter the required change to the transmitter output level. Select Next after the correction.
  7. In the next view, select Apply to take the adjustment into use. If you do not want to save the adjustment, select Back.
  8. Remove the multimeter, close the transmitter cover, and reattach the probe body.