Figure 1. Analog output level adjustment in Insight PC software
1
Enable the adjustment mode with this selection.
2
Start sending a low level test output (4 mA) with this selection. Enter the value
measured with the multimeter in the Measured value, low point field.
3
Start sending a high level test output (20 mA) with this selection. Enter the value
measured with the multimeter in the Measured value, high point field.
4
After adjusting both low and high points, activate the adjustment with this
selection.
5
The corrections made to the output level are shown here. Enter information about the
calibration and the calibration date.
Detach the probe body and open the transmitter cover.
Connect a multimeter to the mA testing points located above the output screw terminals
on the transmitter component board (see Figure 1).