- 1
- Clasp nut (27 mm hex nut)
- 2
- Fitting body (24 mm hex head): ISO1/2 solid structure or NPT1/2 solid structure
- 3
- Sealing material (use one of the following options):
- LOCTITE® No 542 + activ. No 7649 (T=−55 … +150 °C (-67 … +302 °F))
- MEGA-PIPE EXTRA No 7188 (T=−55 … +170 °C (-67 … +338 °F))
- PTFE tape (T=−60 … +210 °C (-76 … +410 °F))
CAUTION Take care not to damage the probe body. A damaged body makes the probe less tight and may prevent it from going through the clasp nut. |
CAUTION In pressurized processes it is essential to tighten the supporting nuts and screws very carefully to prevent loosening of the probe by the action of pressure. |
CAUTION When HMP378 is installed in a process with a pressure differing from normal atmospheric pressure, you must enter the pressure value of the process as an environmental compensation in order to receive accurate measurement results. For information on entering environmental compensations with the transmitter display interface or Vaisala Insight PC software, see Environmental compensation and measurement settings overview. |