HMT370EX has 2 scalable 4 – 20 mA analog
output channels. Each output has the following configuration options:
- Output parameter selection
- Output scale low end and high end
- Error output level and clipping limits
- Output testing (force output to a specified level for multimeter testing)
Note that the clipping and error limit configuration can only be carried out with Insight PC software. |
The configuration of the outputs can be changed with the transmitter display interface or by connecting the transmitter to Vaisala Insight PC software.