The certification requirements in different regions (for example, ATEX certification in the EU and CML certification in Japan) involve differences in, for example, the specific conditions of use that are applicable to the product in each region. The safety information specific to each region is available in the following chapters:
- ATEX/IECEx safety certification information
- Japan safety certification information
- China safety certification information
- Korea safety certification information
- US (FM) safety certification information
- US/CAN (MET) safety certification information
- UK (UKEX) safety certification information
- CAN (CSA) safety certification information
HMT370EX Series Ex classification by region
The following table shows the Ex classification of HMT370EX in each certification region. For the most up-to-date list of regional certifications, see the HMT370EX product page www.vaisala.com/HMT370EX.
Property | Description/value |
---|---|
NOTE: The Ex classifications show the highest level of compliance. Although lower compliance levels are not shown in the classification, they are also included in the classification. For example, compliance with Division 1 also means compliance with Division 2, and compliance with Zone 0 also means compliance with Zone 1 and Zone 2. | |
Europe (ATEX) | |
Gas classification EU (2014/34/EU) |
II 1 G Ex ia IIC T4 Ga |
Dust classification |
II 1 D Ex ia IIIC T200 85 °C Da |
Safety factors |
Ui = 28 VDC, Ii = 100 mA, Ci = 12.1 nF, Pi = 700 mW, Li = 16 μH |
Environmental specifications | |
Tamb | -40 … +60 °C (-40 … +140 °F) |
Pamb | 0.8 … 1.1 bar |
International (IECEx) | |
Gas classification |
Ex ia IIC T4 Ga |
Dust classification | II 1 D Ex ia IIIC T200 85 °C Da |
Safety factors |
Ui = 28 VDC, Ii = 100 mA, Ci = 12.1 nF, Pi = 700 mW, Li = 16 μH |
Environmental specifications | |
Tamb | -40 … +60 °C (-40 … +140 °F) |
Pamb | 0.8 … 1.1 bar |
Japan (CML) | |
Ex classification |
Ex ia IIC T4 Ga Ex ia IIIC T200 85°C Da CML 21JPN2417X |
China (NEPSI) | |
Ex classification |
Ex ia IIC T3~T6 Ga GYJ21.1325X |
Korea (KCs) | |
Ex classification |
Ex ia IIC T4 Ga Ex ia IIIC T200 85 °C Da -40 °C ≤ Tamb ≤ +60 °C IECEx EESF 20.0044.X 21-KA4BO-0891X, 21-KA4BO-0892X |
US (FM) | |
Ex classification |
Class I, Zone 0, AEx ia IIC T4 Ga Zone 20, AEx ia IIIC T85°C Da IS Class I, Division 1, Groups A, B, C, and D T4 IS Class II, III, Division 1, Groups E, F, and G T85°C |
US/CAN (MET) | |
Ex classification |
Class I, Zone 0, AEx ia IIC T4 Ga Class II, Zone 20, AEx ia IIIC T 85°C Da Class I, Division 1, Groups A, B, C, and D T4 Class II, Division 1, Groups E, F, and G T 85 °C Ex ia IIC T4 Ga Ex ia IIIC T 85 °C Da |
UK (UKEX) | |
Ex classification |
II 1 G Ex ia IIC T4 Ga II 1 D Ex ia IIIC T200 85 °C Da −40 °C ≤ Tamb ≤ +60 °C CML 21UKEX2316X |
Canada (CSA) | |
Ex classification |
Ex ia IIC T4 Ga Ex ia IIIC T85°C Da IS Class I, Division 1, Groups A, B, C, and D T4 Class II, III, Division 1, Groups E, F, and G |