Measurement type | Probe models |
---|---|
Humidity and temperature | HMP1, HMP3, HMP4, HMP5, HMP7, HMP8, HMP9 |
Temperature | TMP1 |
Dew point | DMP5, DMP6, DMP7, DMP8 |
Carbon dioxide | GMP251, GMP252 |
Vaporized hydrogen peroxide | HPP271, HPP272 |
Moisture in oil | MMP8 |
Device or series | Models |
---|---|
MHT410 Moisture, Hydrogen and Temperature Transmitter | MHT410 |
Indigo80 Handheld Indicator | Indigo80 |
Device or series | Models |
---|---|
MHT410 Moisture, Hydrogen and Temperature Transmitter | MHT410 |
Polaris™ Process Refractometers 1 | PR53AC, PR53AP, PR53GC, PR53GP, PR53M, PR53SD, PR53W |
Indigo80 Handheld Indicator | Indigo80 |
Property | Description/Value |
---|---|
For use in wet locations | Yes |
Operating humidity |
0 … 100 %RH |
Maximum operating altitude |
4000 m (13 123 ft) |
IP rating |
IP66: Dust-tight. Protected from powerful water jets from any direction. 2 |
Operating temperature | |
With display | −20 … +60 °C (−4 … +140 °F) |
Without display | −40 … +60 °C (−40 … +140 °F) |
Storage temperature | |
With display |
−30 … +60 °C (−22 … +158 °F) |
Without display |
−40 … +60 °C (−40 … +140 °F) |
Property | Description/Value |
---|---|
Operating power | |
Protective extra-low voltage (PELV) |
11 … 35 V DC, 24 V AC ±15 % 50/60 Hz, max. current 2 A (power supply is galvanically isolated) Fuse size for power supply: 3 A Isolation voltage: 500 V AC, 1000 V DC |
PELV power cable temp. rating | ≥ +80 °C (+176 °F) |
Typical current consumption at +20 °C (+68 °F) (Uin 24 V DC) 3 | |
Base consumption (no display, analog outputs, or communication) | 50 mA |
With display | + 60 mA |
With voltage analog output | < 2 mA per channel |
With current analog output | + 21 mA per channel |
With Ethernet cable connected | + 15 mA |
Analog outputs | |
Number of analog outputs | 2 |
Isolation | Isolated from power supply |
Selectable voltage output types | 0 … 1 V, 0 … 5 V, 0 … 10 V, scalable |
Selectable current output types | 4 … 20 mA, 0 … 20 mA, scalable |
Max. wire size | 2.5 mm2 (14 AWG) |
Accuracy of analog outputs at +20 °C (+68 °F) | ±0.05 % full scale |
Temperature dependence | ±0.005 % / °C full scale |
External loads: | |
Current outputs | RL < 500 Ω |
0 … 1 V output | RL > 2 kΩ |
0 … 5 V and 0 … 10 V outputs | RL > 10 kΩ |
Ethernet interface | |
Supported standards | 10BASE-T, 100BASE-TX |
Connector | 8P8C (RJ45) |
Supported protocols | Modbus TCP/IP (port 502), HTTPS (port 8443) |
Property | Description/Value |
---|---|
EU directives and regulations |
EMC Directive (2014/30/EU) RoHS Directive (2011/65/EU) amended by 2015/863 |
Electromagnetic compatibility (EMC) |
IEC/EN 61326‑1, industrial environment CISPR 32 / EN 55032, Class B |
Electrical safety |
IEC/EN 61010‑1 |
Type approvals |
DNV GL certificate no. TAA000032M |
Compliance marks |
CE, China RoHS, FCC, RCM, UKCA |
Listing marks |
UL Listed (USA and Canada) |
FCC compliance |
FCC Part 15, Class B |
Property | Description/Value |
---|---|
For use in wet locations | Yes |
Operating humidity |
0 … 100 %RH |
Maximum operating altitude, AC (mains) power |
3000 m (9843 ft) |
Maximum operating altitude, protective extra-low voltage (PELV) and Power over Ethernet (PoE) | 4000 m (13 123 ft) |
IP rating |
IP66: Dust-tight. Protected from powerful water jets from any direction. 4 |
Operating temperature | |
With display | −20 … +55 °C (−4 … +131 °F) |
Without display | −40 … +60 °C (−40 … +140 °F) |
Without display with barometer module | −40 … +55 °C (−40 … +131 °F) |
Storage temperature | |
With display |
−30 … +60 °C (−22 … +158 °F) |
Without display |
−40 … +60 °C (−40 … +140 °F) |
Property | Description/Value |
---|---|
Operating power 5 | |
Protective extra-low voltage (PELV) version | 15 … 35 V DC, 24 V AC ±20 % 50/60 Hz, max. current 2 A (power supply is
galvanically isolated) Fuse size for power supply: 3 A Isolation voltage: 500 V AC, 1000 V DC |
PELV power cable temp. rating | ≥ +80 °C (+176 °F) |
AC (mains) power version |
100 … 240 V AC 50/60 Hz, max. current 1 A (power supply is galvanically isolated) Fuse size for power supply: 10 A Isolation voltage: 1500 V AC |
Power over Ethernet version 6 |
Power over Ethernet (PoE) IEEE 802.3at Type 2 Class 4 Max. current 600 mA, max. power consumption 25.5 W Isolation voltage: 500 V AC, 1000 V DC |
Typical current consumption at +20 °C (+68 °F) (Uin 24 V DC) 7 | |
Base consumption (no display, analog outputs, or communication) | 50 mA |
With display | + 60 mA |
With voltage analog output | < 2 mA per channel |
With current analog output | + 21 mA per channel |
With relays | + 9 mA per relay |
With Ethernet cable connected | + 15 mA |
With barometer module | + 5 mA |
Analog input | |
Available ranges | 4 ... 20 mA |
Resolution | 6 μA |
Display resolution | 0.01 mA |
Accuracy | ±0.05 mA |
Input impedances | 200 Ω |
Isolation | Isolated from power supply |
Overload protection | 40 mA max. (reverse polarity protected) |
Analog outputs | |
Number of analog outputs | 4 |
Isolation | Isolated from power supply |
Selectable voltage output types | 0 … 1 V, 0 … 5 V, 0 … 10 V, scalable |
Selectable current output types | 4 … 20 mA, 0 … 20 mA, scalable |
Max. wire size | 2.5 mm2 (14 AWG) |
Accuracy of analog outputs at +20 °C (+68 °F) | ±0.05 % full scale |
Temperature dependence | ±0.005 % / °C full scale |
External loads: | |
Current outputs | RL < 500 Ω |
0 … 1 V output | RL > 2 kΩ |
0 … 5 V and 0 … 10 V outputs | RL > 10 kΩ |
Relay outputs | |
Number and type of relays | 2 pcs, SPDT |
Max. switching power, current, voltage |
30 W, 1 A, 40 V DC / 28 V AC |
Max. wire size in PELV version | 2.5 mm2 (14 AWG) |
Max. wire size in AC (mains) version | 1.5 mm2 (16 AWG) |
Ethernet interface | |
Supported standards | 10BASE-T, 100BASE-TX |
Connector | 8P8C (RJ45) |
Supported protocols | Modbus TCP/IP (port 502), HTTPS (port 8443) |
Property | Description/Value |
---|---|
EU directives and regulations |
EMC Directive (2014/30/EU) Low Voltage Directive (2014/35/EU) RoHS Directive (2011/65/EU) amended by 2015/863 |
Electromagnetic compatibility (EMC) |
IEC/EN 61326‑1, industrial environment CISPR 32 / EN 55032, Class B |
Electrical safety |
IEC/EN 61010‑1 |
Type approvals |
DNV GL certificate no. TAA000032M |
Compliance marks |
CE, China RoHS, FCC, RCM, UKCA |
Listing marks |
UL Listed (USA and Canada) |
FCC compliance |
FCC Part 15, Class B |
Property | Description/Value |
---|---|
UL 50E (NEMA) rating |
NEMA 4: Hose down and splashing water, NEMA 250-2014 Type 4 |
Housing classification | IK08, DIN EN ISO 11997-1: Cycle B (VDA 621-415) |
Housing material | AlSi10Mg (DIN 1725) |
Display window material | Strengthened glass (IK08) |
Weight | 1.5 kg (3.3 lb) |
Dimensions (H × W × D) | 142 × 182 × 67 mm (5.63 × 7.17 × 2.64 in) |
Cable diameters for cable glands | |
M20×1.5 glands | 5.0 … 9.0 mm (0.20 … 0.31 in) |
M20×1.5 glands with split bushing | 7 mm (0.28 in) |
M16×1.5 glands | 2.0 … 6.0 mm (0.08 … 0.24 in) |
Property | Description/Value |
---|---|
User interfaces | Web interface for remote use, optional touchscreen display |
Supported languages | English, Chinese, French, German, Japanese, Spanish |
Optional display | 5" capacitive touchscreen |
Integrated data logging capabilities | Non-volatile memory, at least 10 years' storage with 24 h interval logging |
CAUTION Do not modify the unit or use it in ways not described in the documentation. Improper modification or use may lead to safety hazards, equipment damage, failure to perform according to specification, decreased equipment lifetime, or the warranty becoming void. |
CAUTION Do not use the transmitter in a manner not specified by Vaisala. If the transmitter is used in an unspecified manner, the protection provided by the equipment may be impaired. |