Indigo500 specifications - INDIGO520 - INDIGO500SERIES - INDIGO510

Indigo510, Indigo520 Quick Guide

Document code
M212290EN
Revision
D
Language
English (United States)
Product
INDIGO520
INDIGO500SERIES
INDIGO510
Document type
User guide
Table 1. Operating environment
Property Description/Value
Operating temperature

Indigo520 with display −20 … +55 °C (−4 … +131 °F)

Indigo510 with display −20 … +60 °C (−4 … +140 °F)

Without display −40 … +60 °C (−40 … +140 °F)1

Storage temperature

With display −30 … +60 °C (−22 … +158 °F)

Without display −40 … +60 °C (−40 … +140 °F)

For use in wet locations Yes
Operating humidity

0 … 100 %RH

Maximum operating altitude

3000 m (9843 ft)

IP rating

IP66: Dust-tight. Protected from powerful water jets from any direction. 2

Table 2. Inputs and outputs
Property Description/Value
Indigo520 operating power
Protective extra-low voltage (PELV) version 3 15 … 35 V DC, 24 V AC ±20 % 50/60 Hz, max. current 2 A

Fuse size for power supply: 3 A

PELV power cable temp. rating ≥ +80 °C (+176 °F)
AC (mains) power version 3 100 … 240 V AC 50/60 Hz, max. current 1 A

Fuse size for power supply: 10 A

Power over Ethernet version 3 50 V DC, 600 mA PoE+, IEEE 802.3 at PD

Fuse size for power supply: 2 A

Indigo510 operating power
Protective extra-low voltage (PELV)

11 … 35 V DC, 24 V AC ±15 % 50/60 Hz, max. current 2 A

PELV power cable temp. rating ≥ +80 °C (+176 °F)
Indigo520 analog outputs
Number of analog outputs 4, galvanically isolated from power supply
Selectable voltage output types 0 … 1 V, 0 … 5 V, 0 … 10 V, scalable
Selectable current output types 4 … 20 mA, 0 … 20 mA, scalable
Max. wire size 2.5 mm2 (14 AWG)
Accuracy of analog outputs at +20 °C (+68 °F) ±0.05 % full scale
Temperature dependence ±0.005 % / °C full scale
External loads:
Current outputs RL < 500 Ω
0 … 1 V output RL > 2 kΩ
0 … 5 V and 0 … 10 V outputs RL > 10 kΩ
Indigo510 analog outputs
Number of analog outputs 2, galvanically isolated from power supply
Selectable voltage output types 0 … 1 V, 0 … 5 V, 0 … 10 V, scalable
Selectable current output types 4 … 20 mA, 0 … 20 mA, scalable
Max. wire size 2.5 mm2 (14 AWG)
Accuracy of analog outputs at +20 °C (+68 °F) ±0.05 % full scale
Temperature dependence ±0.005 % / °C full scale
External loads:
Current outputs RL < 500 Ω
0 … 1 V output RL > 2 kΩ
0 … 5 V and 0 … 10 V outputs RL > 10 kΩ
Indigo520 relay outputs
Number and type of relays 2 pcs, SPDT
Max. switching power, current, voltage

30 W, 1 A, 40 V DC / 28 V AC

Max. wire size in PELV version 2.5 mm2 (14 AWG)
Max. wire size in AC (mains) version 1.5 mm2 (16 AWG)
Ethernet interface
Supported standards 10BASE-T, 100BASE-TX
Connector 8P8C (RJ45)
Supported protocols Modbus TCP/IP (port 502), HTTPS (port 8443)
Table 3. Compliance
Property Description/Value
Indigo520 EU directives

EMC Directive (2014/30/EU)

Low Voltage Directive (2014/35/EU)

RoHS Directive (2011/65/EU) amended by 2015/863

Indigo510 EU directives

EMC Directive (2014/30/EU)

RoHS Directive (2011/65/EU) amended by 2015/863

EMC compatibility

IEC/EN 61326-1, industrial environment

CISPR 32 / EN 55032, Class B

Electrical safety

IEC/EN 61010-1

Indigo520 compliance marks

CE, China RoHS, FCC, RCM

Indigo510 compliance marks

CE, China RoHS, FCC, RCM, UKCA

Listing marks Canada (SGS), US (SGS)
FCC compliance

FCC Part 15, Class B

Table 4. Mechanical specifications
Property Description/Value

NEMA rating

NEMA 4: Hose down and splashing water, NEMA 250-2014 Type 4

Housing classification IK08, DIN EN ISO 11997-1: Cycle B (VDA 621-415)
Housing material AlSi10Mg (DIN 1725)
Display window material Strengthened glass (IK08)
Weight 1.5 kg (3.3 lb)
Dimensions (H × W × D) 142 × 182 × 67 mm (5.63 × 7.17 × 2.64 in)
Cable diameters for cable glands
M20×1.5 glands 5.0 … 8.0 mm (0.20 … 0.31 in)
M20×1.5 glands with split bushing 7 mm (0.28 in)
M16×1.5 glands 2.0 … 6.0 mm (0.08 … 0.24 in)
CAUTION Do not modify the unit or use it in ways not described in the documentation. Improper modification or use may lead to safety hazards, equipment damage, failure to perform according to specification, decreased equipment lifetime, or the warranty becoming void.
CAUTION Do not use the transmitter in a manner not specified by Vaisala. If the transmitter is used in an unspecified manner, the protection provided by the equipment may be impaired.
1 Operating temperature without display with barometer module −40 … +55 °C (−40 … +131 °F).
2 Evaluated by Eurofins, not by UL.
3 The power supply option is selected when ordering the transmitter.