To test that the analog output of the
transmitter is accurate, you can set the probe to output a fixed level of current and view the
reading to verify the output.Figure 1. Analog outputs test mode in web interface
Connect to the web interface.
Log in to the web interface as an administrator user.
Select Inputs and outputs > Analog outputs.
In Output mode, select the current or voltage range for the outputs.
In the Test mode tab, configure test mode settings.
In Test mode, set the test mode on for the analog output you want to test.
In Force output to, enter the current level that you want the transmitter to output or choose
Output error to use the transmitter's error state output level.
Select Apply.
Read the analog output and verify that the output is correct.
After verifying the output, go back to Inputs and outputs > Analog outputs > Test mode and turn the test mode off.
Repeat 5 … 8 for all the analog output channels you want to test.