Configuring analog outputs in web interface - INDIGO500SERIES - INDIGO510 - INDIGO520

Indigo510, Indigo520 User Guide

Document code
M212287EN
Revision
P
Language
English
Product
INDIGO500SERIES
INDIGO510
INDIGO520
Document type
User guide
You need Administrator user rights for this task.
Figure 1. Analog outputs configuration menu in web interface
  1. Connect to the web interface.
  2. Log in to the web interface as an administrator user.
  3. Select Inputs and outputs > Analog outputs.
  4. In Output mode, select the current or voltage range for the outputs.
  5. In the tabs Output 1 … 4, select the output channel you want to configure.
  6. Configure the analog output channel settings:
    1. In Output off/on, set the output channel off to edit settings.
    2. In Source, select the source for the measurement parameter that controls the output channel.
    3. In Parameter, select the measurement parameter that controls the output channel.
    4. In Scale low end and Scale high end, define the lower and upper ends of the measurement scale for the chosen output parameter.
    5. In Error output, define the output level indicating that the device is in error state.
    6. In Low clipping limit, define the point (in mA below the output range) when the output is clipped due to being out of range.
    7. In High clipping limit, define the point (in mA or V above the output range) when the output is clipped due to being out of range.
    8. Select Apply.
  7. Repeat 56 for all the analog output channels you want to use.